flow technology Vögtlin Instruments AG
  Liquid Flow Flow Gas Pressure Systems & Applications Vögtlin Instruments AG
background
Company
Products
Liquid Flow
Gas Flow
Pressure
Systems
Service & Support
News
Contact
TASIGroup
 
         
  Products > Gas Flow > Mass Flow Meters & Controllers > red-y smart series
         
  Mass Flow Meters and Controllers 'red-y smart series'
         
  INSTRUMENT CHARACTERISTICS
Profibus red-y smart controller
– Thermal mass measurement principle
– Accuracy: 0,3% of full scale + 0,5% of reading
– Turndown ratio: 1:100
– Response time: 50ms
– Communication with PC and SPS
Profibus-Connection
– Real gas calibration
– Insensitive to pressure and temperature changes
– Minimum pressure loss
– Materials: Aluminium or stainless steel
– Temperature measurement as standard (digital)
– Compact design, simple installation
– Easy to service and maintain
– 3-year warranty
– Multigas - One device for 10 different gases
  Instrument versions:
  standard: The economic solution
(Accuracy ± 1.0% of full scale, Turndown ratio 1 : 50)

hi-performance: With highest accuracy and turndown ratio
(Accuracy ± 0.3% of full scale and ± 0.5% of reading, Turndown ratio 1 : 100)
   
  More Information:
  » Instrument Types
» Technical Overview
» NEW: free software get red-y 5
» Accessories
» Applications & Solutions
» Online Quotation
» Troubleshooting
» Documentation, Downloads
» Contact
   
 
  INSTRUMENT TYPES
  mass flow meter red-y smart meter red-y smart meter GSM
Thermal mass flow meter
  Mass flow controller red-y smart controller red-y smart controller GSC
Thermal mass flow controller
  mass flow controllers for OEM OEM version
For customer-specific requirements
   
 
  TECHNICAL OVERVIEW
  Performance data Media Air, O2, N2, He, Ar, CO2, H2, CH4, C3H
Other gases and gas mixtures on request
    Response time 50 ms
    Repeatability ± 0.2% of full scale
    Longterm stability < 1% of measured value / year
    Power supply 24 Vdc (18 – 30 Vdc), 15 Vdc on request
    Current consumption Meter: max. 100mA
Controller: max. 250mA
    Operation pressure 0.2 – 11 bar a
    Temperature 0 – 50°C
    Materials Anodized aluminium, optional stainless steel electropolished
    Seals FKM, optional EPDM
    Pressure sensitivity < 0.2% / bar of reading (typical N2)
    Temperature sensitivity < 0.025% FS measuring range type / °C
   
  Measuring Ranges Type Measuring range (Air), full scale freely selectable
  smart meter GSM: GSM-A from 0 ... 25 mln/min to 0 ... 600 mln/min
    GSM-B from 0 ... 600 mln/min to 0 ... 6000 mln/min
    GSM-C from 0 ... 6 ln/min to 0 ... 60 ln/min
    GSM-D from 0 ... 60 ln/min to 0 ... 450 ln/min
  smart controller GSC: GSC-A from 0 ... 25 mln/min to 0 ... 600 mln/min
    GSC-B from 0 ... 600 mln/min to 0 ... 6000 mln/min
    GSC-C from 0 ... 6 ln/min to 0 ... 60 ln/min
    GSC-D from 0 ... 60 ln/min to 0 ... 450 ln/min
   
  Integration Output signals:  
    - Analog 0..20 mA, 4..20 mA, 0..5 V, 1..5 V, 0..10 V, 2..10 V
    - Digital RS-485; Modbus RTU (Slave); Lab View-VI‘s available Optional: ProfiBus DP-V0, DP-V1
    Process connection G¼“ female less than 60 ln/min, G½“ female less than 450 ln/min
    Inlet section None required
    Electrical connection Sub D plug, 9 pole
    Mounting orientation Any orientation (horizontal only above 5 bar)
   
  Safety Test pressure 16 bar a
    Leak rate < 1 x 10-6 mbar l/s He
    Environ. protection IP-50
    EMC EN 61326-1
   
 
  ACCESSORIES
  Software get red-y Free Software 'get red-y'
– Device information (flow rate and temperature)
– Operating state monitoring
– Switch measurement units & gas type
– Plug & play with USB cable
– Datalogger
– Calibration & adjustment
– Runs on all computers with Windows XP, Vista, 7
» More Information
   
  Display and Control Device PCU-10 Display and Control Device PCU-10
– Display of the gas flow set value for each connected meter
   and controller
– Display of the gas flow actual value for each connected meter
   and controller
– Specification and storage of set values
– Storage of up to 12 different gas mixtures as recipes
– Total flow meter for each connected meter and controller
– Specification of minimum and maximum limit values with red alarm
   display on reaching the limits
– Gas mixer with up to 12 recipes
– Display of the instrument temperature of the connected meter
   and controller
– Mixed operation with Vögtlin and BHT devices possible
– No additional mains adaptor required
– Control of the connected meter and controller with a PC by means
   of PDM-U connection cable
– Use of the get red-y software on a PC for direct configuration
   of the meter and controller
» More Information: www.red-y.com
   
    Cables & Modules
Cables and Modules for the Thermal Mass Flow Meters and Mass Flow Controllers 'red-y smart series'.
» More Information: www.red-y.com
   
 
  DOWNLOADS
  Document Type/Size
  Data Sheet red-y smart series PDF/750kB
  Data Sheet Software 'get-red-y' PDF/2.5MB
  Data Sheet Readout System PCU-10 PDF/500kB
  Data Sheet Cables, Modules & Supply PDF/100kB
  Data Sheet PROFIBUS PDF/150kB
  Data Sheet Fittings PDF/300kB
  CAD 3D models  
   
  » Flow Solutions and Applications  
  » More Downloads  
   
 
  Vögtlin Instruments AG – flow technology – Phone +41 61 756 63 00 – Last Update: 13.01.2012